February 28, 2019
Introduction
These release notes describe how to apply this update, and the changes that have been made since the previous firmware release.
Applying this update will not alter the device setup, or clear log files stored on external media.
[Top]
Updates are applied using FTP, SFTP, or SCP. When using SFTP or SCP, connect to port 22, use "HCI" as the user name, and by default there is no password.
While applying this update, take care not to interrupt the power to the instrument.
Perform the following steps to update your icListen:
- Power the Host Interface using the supplied power adapter.
- Connect to the Host Interface via FTP, SFTP or SCP.
- Once connected, you should see a directory called "update". If this directory does not exist, create it. This directory name is case sensitive.
- Upload the update ".icu" file into the "update" directory.
- The instrument must now be reset in order for the update to take place. This can be done through the web interface, or using Ocean Sonics' Marco software.
- Marco: Use the "Reset" button on the lower left area main display.
- Web Interface: Type the IP address of your Host Interface into the address bar of your web browser (Firefox, Internet Explorer, etc), and use the "Reset" button on the "Power Operations" page to reset the instrument.
- Allow approximately 5 minutes for the update to complete.
- The update should now be complete. You may connect using the web browser to ensure the update has successfully taken place. The "Firmware Release" number should be displayed on the "Home" page. If the release version is blank, or does not appear to be correct, try refreshing the page in your browser.
[Top]
Changes since previous major release:
- Enhancements
- Rebuid operating system for improved efficiency and security.
- Bug Fixes
- Fix bug which in some circumstances could cause files to be recorded in the root folder instead of data folder.
- Fix bug where returned setup type for unrecognized setup responses indicated the desired setup type instead of the unrecognized one.
- Fix bug where job setup response payload was offset by 1 byte.
- Fix bug which would fail to report invalid waveform log file sizes when logging on hydrophones.
- Fix bug where invalid FFT log length setups caused invalid FFT log mode errors to be reported.
- Fix bug which could cause system to become unresponsive in response to invalid command and control messages being received.
- Fix bug which could cause command and control responses to become corrupted if new command and control messages were received before the response was fully sent.
- Version updates
- Firmware upgraded to v1.1.00
- Web Interface upgraded to v1.1.00
- Polo upgraded to v1.3.04
[Top]
1 - Hardware release 2 introduces buzzer functions, battery monitoring, and PPS synchronization. These features are not available in hardware release 1.